Recall Z-1611-2008 ยท IXR Table, Radiographic, Tilting

GE Healthcare

Advantx Legacy Radiographic and Fluoroscopic systems, model 2102424, 2118104, 2157009, 2183000, 2206038, 2206048, 2206056, 2224559, 2258627, 2268730, 2330340, 2333974, 2118104-2, 2118104-3, 2118104-4, 2118104-5, 2118104-6, 2118104-7, 2118104-8, 2139574-2, 2147592-13, 2171204-2, 46-262751G6, 46-262751G7, 46-262751G8, 46-275382G58, and 46-302056G1. GE Medical Systems - Americas: Milwaukee, USA. T

DateSep 19, 2008 (initiated Jan 2, 2008)
StatusTerminated
ReasonGE Healthcare was notified that 2 recently installed Precision 500D systems were not in compliance with portions of the Performance Standard for Diagnostic X-ray Equipment. X-ray production was possible from the fluoroscopic X-ray tube when the Primary Protective Barrier was not in position to intercept the X-ray beam. Users are recommended to follow good clinical standards to ensure the Image I
Root causeNonconforming Material/Component
ActionA Product Safety Notification letter, dated 01/09/2008 was sent to affected customers. This letter described the safety issue, affected product and safety instructions. A GE Field Engineer will schedule a field visit to affected customers to inspect and replace the park latch switch as needed.
Quantity2863 total (2744 US / 119 OUS)
DistributionNationwide, Washington D.C. and Puerto Rico and OUS to include: ARGENTINA, AUSTRALIA, BRAZIL, CANADA, CHILE, DOMINICAN REPUBLIC, EGYPT, GERMANY, HONDURAS, INDIA, INDONESIA, IRELAND, JAMAICA, KOREA, KUWAIT, MALAYSIA, MEXICO, NETHERLANDS, PANAMA, RUSSIAN FEDERATION, SAUDIA ARABIA, SYRIAN ARAB REPUBLI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973039LEGACY/LEGACY-D TABLE
GE Medical Systems, Oct 16, 1997