Recall Z-1610-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare

GE Precision 500D, model 2288798, 2289299, 2290479, 2305472, 2305473, 2305495, 2336900, 2345243, 2401181, 2403790, 2403791, 2404103, 2407276, 2407576, 5179385, 2289299-2, 2290479-2, 2345260-12, 2345260-3, 2403790-3, and 2403791-3, GE Medical Systems - Americas: Milwaukee, USA. The device is used to perform fluoroscopic examination of human anatomy.

DateSep 19, 2008 (initiated Jan 2, 2008)
StatusTerminated
ReasonGE Healthcare was notified that 2 recently installed Precision 500D systems were not in compliance with portions of the Performance Standard for Diagnostic X-ray Equipment. X-ray production was possible from the fluoroscopic X-ray tube when the Primary Protective Barrier was not in position to intercept the X-ray beam. Users are recommended to follow good clinical standards to ensure the Image I
Root causeNonconforming Material/Component
ActionA Product Safety Notification letter, dated 01/09/2008 was sent to affected customers. This letter described the safety issue, affected product and safety instructions. A GE Field Engineer will schedule a field visit to affected customers to inspect and replace the park latch switch as needed.
Quantity2863 total (2744 US / 119 OUS)
DistributionNationwide, Washington D.C. and Puerto Rico and OUS to include: ARGENTINA, AUSTRALIA, BRAZIL, CANADA, CHILE, DOMINICAN REPUBLIC, EGYPT, GERMANY, HONDURAS, INDIA, INDONESIA, IRELAND, JAMAICA, KOREA, KUWAIT, MALAYSIA, MEXICO, NETHERLANDS, PANAMA, RUSSIAN FEDERATION, SAUDIA ARABIA, SYRIAN ARAB REPUBLI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011624EXPEDIO 500D R&F X-RAY SYSTEM
GE Medical Systems, Jun 8, 2001