Recall Z-0112-2009 ยท HGM System, Monitoring, Perinatal

GE Healthcare Integrated IT Solutions

Centricity Perinatal (formerly Quantitative Sentinel) System - Fluid Total Precision software; automatic patient data management providing clinical information at the bedside in Labor & Delivery, Mother-Baby and the Neonatal Intensive Care Unit; GE Healthcare Integrated IT Solutions, Barrington, IL 60010

DateOct 15, 2008 (initiated Sep 25, 2008)
StatusTerminated
ReasonOn the I&O chart, the IN, OUT and NET fluid totals values will not honor the numeric precision configuration, always displaying a whole number, losing decimal point accuracy.
Root causeSoftware design
ActionGE Healthcare notified their Centricity Perinatal customers via letter dated 4/20/07 of the problem with the Fluid I & O Module when fluid precision is required. The letter instructed those accounts that require fluid precision to contact GE Technical Support at 1-800-433-2009 to obtain a software patch to fix the problem. A follow-up letter dated 8/29/08 was sent on the same date to the accounts informing them that GE had re-classified the issue as a class II safety issue and that all customers need to have their software updated. The accounts were requested to contact GE Tech Support at 8
Quantity470 units
DistributionWorldwide Distribution --- including USA and country of Canada.