Recall Z-0460-2009 ยท LLZ System, Image Processing, Radiological
GE Healthcare Integrated IT Solutions
GE Centricity PACS RA1000 Workstation (for diagnostic image analysis). GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity" PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals. It is also intended for use as a clinical review workstation throughout the healthcare facility. The workstation
| Date | Dec 13, 2008 (initiated Sep 25, 2008) |
| Status | Terminated |
| Reason | Software anomalies result in patient safety issues involving patient jacket content intermittently becoming unintentionally out of synchronization with the images being displayed, and involving Default Display Protocols (DDPs) which are used to layout images when displaying a study. If the user does not check the Study Date Time of the exam on the image title bar, then they may interpret the curr |
| Root cause | Software design |
| Action | GE Healthcare Integrated IT Solutions notified their Centricity PACS RA1000 Workstation customers via letter (Product Safety Notification) dated 9/25/08 of the software anomalies resulting in patient safety issues involving the patient jacket content intermittently becoming unintentionally out of synchronization with the images being displayed, and involving Default Display Protocols (DDPs) which are used to layout images when displaying a study. If the user does not check the Study Date Time of the exam on the image title bar, then they may interpret the current exam as a historical study and |
| Quantity | 546 units |
| Distribution | Worldwide Distribution including USA and countries of Australia, Belgium, Brazil, Canada, Cayman Islands, Chile, Denmark, Ecuador, France, Germany, Hong Kong, Italy, Japan, Korea, Malaysia, Malta, Mexico, New Zealand, Singapore, Spain, Switzerland, Taiwan, United Kingdom and Venezuela. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K043415 | CENTRICITY PACS SYSTEM Ge Medical Systems Information Technologies, Jan 21, 2005 |