Recall Z-0187-2009 ยท MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor

Medtronic Neuromodulation

Medtronic DBS" 3389 Lead Kit for Deep brain stimulation, DBS Leads: Proximal 40 mm, Distal 9.0 mm, Medtronic, Inc. Medtronic Active Therapy includes Activa Parkinson's Control Therapy and Activa Tremor Control Therapy. Parkinson's Control Therapy: Bilateral stimulation of the internal globus pallidus (GP) or the subthalamic nucleus (STN) using Medtronic Activa Parkinson's Control Therapy is indi

DateOct 27, 2008 (initiated Jun 14, 2008)
StatusTerminated
ReasonMedtronic has received reports of DBS" leads being damaged at the proximal connector end of the lead when the lead cap is used in the implant procedure. The lead damage is the result of excessive torsion and / or tensile lading of the proximal connector end of the lead. Data also suggest that excessive torque while tightening or loosening the setscrews may twist the setscrew connector block and d
Root causeProcess design
ActionConsignees were sent a "Medtronic Urgent: Medical Device Correction" letter 8/2008. The letter addressed to Healthcare Professionals, described the product involved, explanation of the issue, recommendations for Implant and included an attachment "Updated Instructions for Using the Lead Cap". Contact Medtronic Neuromodulation at 1-800-707-0933 for assistance.
Quantity9494
DistributionAL, AZ, AR, CA, CO, CT, FL, DC, GA, HI, ID, IL , IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE , NV, NH, NJ, NM, NY, NC, ND, OH, OR, OK, PA, RI, SC, SD, TN, TX, UT , VT, VA, WA, WV, and WI. ALGERIA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CROATIA, CYPRESS, CZECH REPUBLIC, DENMARK, EGY