ClearanceBrief
Recall Z-0026-2009 ยท HET Laparoscope, Gynecologic (And Accessories)

Applied Medical Resources Corp

Applied Medical, Separator Abdominal Access System REF: C0607 15x150mm Threaded Separator System with Universal Seal, for laparoscopic procedures.

DateOct 7, 2008 (initiated Jun 25, 2008)
StatusTerminated
ReasonPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Root causeProcess control
ActionCustomers were notified via FedEx or UPS on June 25,2008 with an overnight letter. The letter advises that Applied Medical is conducting a voluntary recall of the 15mm Separator Access Systems, Models C0604, C0605, C0606 and C0607 due to the potential inability to insufflate through the stopcock. Per the letter, the firm is asking that all 15mm product listed in the letter be returned. A response form is provided.
Quantity14,113 total
DistributionNationwide and Canada distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012884DILATING TIP OBTURATOR
Applied Medical Resources Corp., Jan 17, 2002
K945741TROCAR AND CANNULA ACCESSORY
Applied Medical Resources, Feb 28, 1995