Recall Z-0087-2009 ยท BSZ Gas-Machine, Anesthesia

Draeger Medical, Inc.

Draeger Medical Apollo Anesthesia Machine, catalog number 8606500

DateOct 8, 2008 (initiated Aug 22, 2008)
StatusTerminated
ReasonSporadic errors in various device functions, including low readings of the single gas flows for oxygen, nitrous oxide, or air delivery, or mechanical ventilation failure. Alarms and error messages function as designed.
Root causeNonconforming Material/Component
ActionDraeger Medical sent a Medical Device Recall letter to their customers via certified mail on 8/22/08 informing them of the problem. Until the exchange of affected printed circuit boards has been implemented, the firm is recommending that users closely monitor all device displays and error messages and if applicable act in accordance with the advisory statements and remedial information provided in the Instructions for Use as well as making provisions for alternative patient care measures.
Quantity214 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042607PRIMUS US
Draeger Medical, Inc., Nov 26, 2004