Recall Z-0175-2009 ยท LFK Catheter, Femoral

ConMed Corporation

VCARE (Vaginal-Cervical-Ahluwalia's-Retractor-Elevator) STANDARD, Uterine Manipulator, REF 60-6085-000. Made in USA, CONMED CORPORATION, UTICA, NEW YORK USA 13502. The ConMed VCARE Retractor/Elevator is indicated for manipulation of the uterus and injection of fluids or gases during laparoscopic procedures such as Laparoscopic Assisted Vaginal Hysterectomy (LAVH), Total Laparoscopic Hysterectom

DateOct 28, 2008 (initiated Sep 8, 2008)
StatusTerminated
ReasonThe firm received several complaints sighting detachment of the balloon at the distal end of the shaft. It was also noted that the incidence of the forward cup slipping off the shaft and being retained in the vaginal canal had increased.
Root causeProcess change control
ActionThe domestic recall notices were mailed on 9/08/08 by USP Priority Mail pouches. The international notices were sent via FedEx International Priority on 9/08/08. The letter described the problem, the health risk and requested that customers inventory the product and return them back to the firm. Customers are requested to fax the reply form to the firm. Contact ConMed Corporation at 1-800-765-8375, ext 3237 for assistance.
Quantity32,453 (total devices)
DistributionNationwide, Australia, Belgium, Brazil, Chile, Ecuador, France, and Switzerland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071907VCARE RETRACTOR/ELEVATOR
Conmed Corporation, Oct 5, 2007