Recall Z-0118-2009 ยท LLZ System, Image Processing, Radiological

Philips Healthcare Informatics, Inc.

iSite PACS (Picture Archiving and Communication System), software versions 3.6.28.x and 4.1.x. The product is an image management system (software package) used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. Device is intended for use by trained professionals.

DateOct 28, 2008 (initiated Aug 29, 2008)
StatusTerminated
ReasonTwo defects have been identified: 1) potential to display a patient on the canvas page that is different from the patient whose images are displayed on diagnostic monitors when using Conference Presentation States and when opening two or more studies from a folder or an exam worklist and 2) potential to miscalculate measurements when Pixel Spacing and Imager Pixel Spacing DICOM tag values are bot
Root causeSoftware design
ActionA notification letter (Urgent Product Correction) dated 8/29/08 was sent via certified mail to all customers alerting them of the issue, requiring posting of notification. Philips Healthcare Informatics, Inc. strongly recommend all customers advise their iSite Radiology users not to open multiple patient exams simultaneously from a folder or worklist using Conference Presentation States. A software correction is being worked on, and should be available by end of Q3 2008 for version 4.1.45.2. A mandatory upgrade is required for all customers who are currently using iSite version 4.1.45.2 a
Quantity27 units.
DistributionWorldwide Distribution including USA (MN, CO and PA) and countries of Italy, Germany, Netherlands, Sweden, France, Portugal, Switzerland, Thailand, Japan and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042292ISITE PACS
Stentor, Inc., Sep 8, 2004
K063267ISITE PACS, V 4. X
Stentor Inc., A Philips Medical Systems Co., Nov 9, 2006