Recall Z-2341-2008 ยท MJK Dna Probe, Trichomonas Vaginalis
Physio Control, Inc.
LIFEPAK CR Plus Automated External Defibrillator. Device is primarily deployed in Public Access Defibrillator (PAD). It is a small lightweight device intended for use by minimally trained responders to treat victims of sudden cardiac arrest... Product is capable of 30 full discharges. Labeling on the device states LIFEPAK CR Plus LIFEPAK EXPRESS Defibrillators with ADAPTIV Biphasic Technology; Me
| Date | Sep 16, 2008 (initiated Aug 28, 2008) |
| Status | Terminated |
| Reason | This recall is being conducted due to the device being configured with the incorrect software for semi-automatic instead of fully automatic use. When the device is needed for a cardiac arrest emergency, the device will require that the user press the shock button instead of automatically delivering a shock as per the normal operation of the fully automatic device. Fully automatic defibrillators h |
| Root cause | Employee error |
| Action | On 8/28/08 the firm began contacting consignees by telephone and stating that the LIFEPAK CR Plus defibrillators were discomfiture as semi-automatic, and that a responder is unable to press the shock button to deliver therapy when instructed by the AED voice prompt because the shock button is covered and not visible. The consignee was instructed that the unit(s) would be replaced immediately. This phone contact was followed by a letter flagged "'URGENT - MEDICAL DEVICE RECALL" dated August 2008 that was f axed or e-mailed the same day. The letter repeated the information provided in the t |
| Quantity | 249 units |
| Distribution | Worldwide Distribution including USA, and countries of Canada, Germany, and Hong Kong. |
| FDA record | Open on fda.gov |