Recall Z-2655-2010 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Sonosite, Inc.

SLT transducer is used on the SonoSite MicroMaxx ultrasound system. The SLT Transducer label is labeled in part: " MicroMaxx SLT...Manufactured for SonoSite, Inc., Bothell, WA 98021".

DateOct 1, 2010 (initiated Nov 2, 2007)
StatusTerminated
ReasonSterilization of the SonoSite SLT ultrasound surgical transducer with the Sterrad NX System may result in damage to the transducer.
Root causeProcess control
ActionOn November 2, 2007, SonoSite sent out the Important Information Regarding Sterrad NX Incompatibility letter and the Technical Bulletin 005/07 Rev A to their consignees. The letter advises the consignees that SonoSite has recently found that sterilization of the SonoSite SLT and LAP surgical ultrasound transducers with the Sterrad NX system may result in damage to the transducer. The Technical Bulletin 005/07 Rev B dated November 2, 2007 informs the consignees that the damage has been observed primarily on the circuit board in the transducer connector, and cable and laparoscope damage may al
Quantity38 units
DistributionDevices were distributed to medical centers and hospitals in AR, CA, IN, KS, MA, MD, MI, NC, NE, NJ, NV, OH, SC, TN, TX, WI, Washington DC and to FOREIGN COUNTRIES Australia, China, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Korea, Republic of, Libyan Arab Jamahiriya, Mexico, Spain, U