Recall Z-0542-2009 ยท JIX Calibrator, Multi-Analyte Mixture

Roche Diagnostics Corp.

Roche ISE Internal Standard Gen.2 for use on Roche cobas c systems, Roche Diagnostics, Inc., Indpls., IN; Catalog no. 04522320190. Product is an in vitro diagnostic.

DateJan 6, 2009 (initiated Sep 15, 2008)
StatusTerminated
ReasonTest results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
Root causePackage design/selection
ActionConsignees were notified by letter dated 9/15/08, which provided a work-around for them to use until the problem can be permanently resolved. The letter was entitled "URGENT MEDICAL DEVICE CORRECTION." For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
Quantity7909
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060373COBAS 6000 SERIES SYSTEM
Roche Diagnostics, Mar 13, 2006