Recall Z-2198-2008 ยท FMI Needle, Hypodermic, Single Lumen

Greiner Bio-One North America, Inc.

Plastic Cannula Holdex Tube Holder, Item number: 450216. It is intended for use in routine specimen withdrawal from specimen collection bags or bottles through needleless cannula ports.

DateSep 30, 2008 (initiated Aug 5, 2008)
StatusTerminated
ReasonAt removal of the tube from the holder the needle may dislodge and become stuck in the tube stopper with the blunt end of the needle facing out.
Root causeProcess design
ActionConsignees were notified by letter on 08/06/2008. Letters to Distributors instructed them to stop distribution of the affected product immediately and to hold any product of the lots they may have pending returned to Greiner bio-one. The were instructed to provide Greiner bio-one with a list of their customers who may have received the affected lot so that they can contact them. A Product Disposition form was included to be completed and returned to Greiner bio-one. Contact Greiner Bio-One North America, Inc. at 1-888-286-3883 for assistance.
Quantity150,000 pcs
DistributionAL, FL, IN, KS, KY, MO, MN, NC, OH, PA, TN, TX, UT, VA, WI and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980768VACUETTE HOLDEX
Greiner Meditech, Inc., Apr 9, 1998