Recall Z-2457-2008 ยท DCK C-Reactive Protein, Antigen, Antiserum, And Control

Cholestech Corp

Cholestech LDX High Sensitivity C-Reactive Protein (hs-CRP) Test Cassette. The Cholestech LDX high sensitivity C-Reactive Protein (hs-CRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum. Measurement of CRP is useful as an aid in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.

DateSep 20, 2008 (initiated Aug 9, 2006)
StatusTerminated
ReasonTest results are high, outside of control material upper limit specification. Results could be as high as 15%.
Root causeDevice Design
ActionCholestech provided its three distributors a notification letter/response form with instructions to distribute the letter to their customers/end users dated August 9, 2006. The letter instructed customers to discontinue use of the lot of cassettes, return remaining cassettes to Cholestech through the Technical Service Representative and to complete and return the response form. The end users returned the signed letters directly to Cholestech so that effectiveness checks could be completed.
Quantity2, 910 cassettes
DistributionNationwide Distribution --- including states of Virginia, Texas and North Carolina, Ohio, Arkansas, Washington, Florida , California and Michigan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040579CHOLETECH LDX HIGH-SENSITIVITY C-REACTIVE PROTEIN (HS-CRP)