Recall Z-2452-2008 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Emotion 6. Computed tomography X-ray system

DateSep 21, 2008 (initiated Aug 5, 2008)
StatusTerminated
ReasonPotential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to open in the event of a short circuit condition between the internal switch contacts.
Root causeNonconforming Material/Component
ActionThe firm issued and Safety Update in July 2008 informing users that the firm will have representatives visit each account to replace the cap per Update Instructions CT032/08/S.
Quantity53 units
DistributionNationwide including states of AK, AL, CO, FL, GA, IA, ID, IL, IN, LA, MD, MN, MS, NC, NE, NY, OH, OK, OR, RI, SD, TX, UT, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023687SOMATOM EMOTION 6 CT SYSTEMS
Siemens Medical Systems, Inc., Nov 22, 2002