Recall Z-1620-2010 ยท LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)

Bio-Rad Laboratories

Bio-Rad's Monolisa (TM) Anti-HBs EIA test kit, Labeled as, "MONOLISA TM Anti-HBs EIA 25220; MONOLISA TM Anti-HBs Calibrator Kit 25219... Bio-Rad Laboratories Redmond, WA 98052" For the qualitative detection of antibody to Hepatitis B Surface Antigen (anti-HBs) in Human Serum and EDTA or citrated plasma.

DateMay 20, 2010 (initiated Aug 11, 2008)
StatusTerminated
ReasonThe package insert provided with this kit contains a typographical error, giving incorrect directions to prepare a Working Conjugate Solution. If these directions are followed, it would result in a 1:6 dilution and invalidate the assay.
Root causePackaging process control
ActionAll domestic customers were notified via an Urgent Product Correction customer letter sent by DHL or FedEx on 8/11/08. Enclosed with the letter was a corrected package insert. The Canadian subsidiary was notified via e-mail on 8/11/08. The letter advised customers of the correct wording, and advised that a new corrected package insert was enclosed and to make note of the correction in the laboratory procedures. Questions should be directed to Bio-Rad Laboratories Technical Support at 1-800-224-6723.
Quantity795 kits
DistributionWorldwide Distribution -- USA and Canada.
FDA recordOpen on fda.gov