Recall Z-0214-2009 ยท JQP Calculator/Data Processing Module, For Clinical Use

Sunquest Information Systems, Inc.

Sunquest Laboratory System. Software Version: Sunquest Laboratory v5.4.2, v6.1, v6.2 and v6.3. The system is marketed as an automated solution for managing laboratory processes including: 1) Patient registration, 2) Order entry and order modification, 3) Specimen collection, verification, suitability, and distribution, 4) Quality assurance checking, 5) Workload recording, 6) Billing charge cap

DateNov 12, 2008 (initiated Aug 1, 2008)
StatusTerminated
ReasonWhile performing in-house testing for the Sunquest Laboratory System and under certain conditions, there have been instances in which an error that occurred at the database level or originated in the operating system (Cache)was not communicated or displayed in the Graphic User Interface (GUI) application so the activity continued. While this error was registered in the error log in the System U
Root causeSoftware design
ActionA notice (PSN-08-07) was issued to all customers on 08/01/2008 via e-mail. The notice advises customers of the problem and actions Sunquest Information Systems has taken to eliminate the Problem. Sunquest Client Support will contact their facility to schedule correcting the error trap setting. The notice also provides temporary procedural instructions to include monitoring error log using System Utilities (Function UTL) and suboption 2, System Error/Message Inquiry. Contact Client Support at 520-570-2252 for questions or additional information.
Quantity139
DistributionWorldwide Distribution including USA and countries of Canada, Denmark, Saudi Arabia and United Arab Emirates.
FDA recordOpen on fda.gov