Recall Z-0285-2009 ยท DSI Detector And Alarm, Arrhythmia

GE Healthcare

GE Healthcare Solar 8000M System, a multiparameter physiological patient monitoring system intended for use on adult, pediatric and neonatal patients, within a hospital or facility environment

DateJan 5, 2009 (initiated Apr 2, 2008)
StatusTerminated
ReasonA software error has been discovered when using Combination Monitoring with Solar 8000M/i - PDM - Telemetry in combo mode. This error has only been noted with this particular configuration and mode of monitoring. It has not been observed in other monitoring applications. The software error will result in a reset of the Solar 8000M/i, causing a loss of monitoring for approximately 1-2 minutes a
Root causeSoftware design
ActionA GE Healthcare consignee letter was sent out 9/9/08. The letter dated August 26, 2008 was addressed to Healthcare Administrator/Risk Manger. The letter described the problem, product and provided recommendation.
Quantity2782
DistributionNationwide to: AL, AR, AZ, CA, FL, GA, IA, ID, IL, IN, MA, MI, MN, MO, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, and WI. OUS to include: Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Egypt, France, Germany, Indonesia, Ireland, Italy, Jordan, Korea, Lebanon, Lithuania, Malaysia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012467SOLAR 8000M SYSTEM
General Electric Medical Systems Information Techn, Feb 11, 2002