Recall Z-0170-2009 ยท IWJ System, Applicator, Radionuclide, Manual

Alpha Omega Services Inc

Alpha-Omega Services, Inc. Product No:NFW0001-010 Description: FLEXIGUIDE NEEDLE, SERIES-15, 25 CM WITHOUT FEMALE LUER ADAPTOR, SHARP AND STYLET WITH 90-DEGREE BEND, 10/PKG Quantity: 10/PKG L-NFW0001-010. The FlexiGuide is a flexible needle designed to provide an enclosed pathway for localized radiation therapy using a Remote High Dose Rate (HDR) Afterloader or manual Low Dose Rate (LDR) method.

DateOct 22, 2008 (initiated Nov 15, 2007)
StatusTerminated
ReasonThere is a possibility that the Flexiguide Needle tip is defective. There is the potential for the needles tip to leak, thereby breaking the sterile boundary and possibly contaminating other devices with body fluids.
Root causeNonconforming Material/Component
ActionNotification of customers initially began November 15, 2007 based on a cursory visual inspection of the defective needle tips. The initial letter advised that there is a possibility that the Flexiguide Needle tip is defective in that there is the potential for the needle's stylet or an HDR source wire/cable to puncture through the defective needle tip during implant or treatment and penetrate the surrounding tissue thereby injuring the patient and/or breaking the sterile boundary. Further distribution and/or use of these products should not take place. The recall notification was updated on N
Quantity40 pkgs (10 each pkg)
DistributionNV, AZ, CA, MI, UT, MO, CO, and OH
FDA recordOpen on fda.gov