Recall Z-2445-2008 ยท HAW Neurological Stereotaxic Instrument

Zimmer Inc.

Universal Optical Tracker Fixation, not sterile, Manufactured by Orthosoft, Montreal, Canada, REF 111.031. Surgical instrument.

DateSep 20, 2008 (initiated Aug 6, 2007)
StatusTerminated
ReasonInadequate weld on tracker, knob oversized and poor thread design on attachment screw results in seizing. The instrument may break during use, resulting in surgical delay and an increased risk of infection.
Root causeDevice Design
ActionFirm's sales representatives were notified via e-mail on 8/6/07 to review their inventory and that of their accounts and to notify the firm of their location so that they could be replaced. Contact the firm by telephone at 574-267-6131 for additional information concerning this recall.
Quantity39
DistributionNationwide Distribution including states of California, Florida and Washington.