Recall Z-2440-2008 ยท JCX Apparatus, Suction, Ward Use, Portable, Ac-Powered

KCI USA, Inc.

500 mL InfoV.A.C. Canister (without Gel); Part #M8275071/5 (5 canister pack) and Part #M8275071/10 (10 canister pack); products are single use and labeled as STERILE; distributed by Kinetic Concepts, Inc., San Antonio, TX.

DateSep 20, 2008 (initiated Jul 29, 2008)
StatusTerminated
ReasonThe port on the 500mL InfoV.A.C. Canister that connects to the suction pump tubing was found to be partially or fully occluded with plastic.
Root causeEquipment maintenance
ActionUrgent - Medical Device Recall letters were sent within the US on 07/29/08 via UPS next day delivery service and a return receipt was maintained to demonstrate delivery of the recall to all consignees. Additionally, firm is sending representatives into accounts which received more than three sales units of canisters and to all foreign consignee locations to facilitate quarantine and replacement of affected stock. Consignees asked to confirm that adequate flow is being delivered and that the dressing is being drawn down for those patients currently using the InfoV.A.C. Canisters. Also, consi
Quantity79,275 canisters.
DistributionWorldwide Distribution --- including USA and countries of Austria, Belgium, Denmark, France, Germany, Netherlands, Sweden, and the UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063740V.A.C. THERAPY SYSTEMS - INFO V.A.C. THERAPY UNIY, MODEL M6257766
Kci USA, Inc., Jun 7, 2007