Recall Z-2438-2008 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

GE Healthcare

Innova 4100 / 4100 IQ , GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector) Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. Also indicated for generating fluoro

DateSep 20, 2008 (initiated Jul 25, 2008)
StatusTerminated
ReasonThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Root causeDevice Design
ActionConsignees (Hospital Administrators; Hospital Risk Management Department; Mangers of Radiology/Cardiology; Radiologists/Cardiologists) were sent a "GE Urgent Medical Device Correction" letter dated July 2008. The letter described the affected products; the problem; and the user instructions, which state that the hospital should establish emergency workflow procedures to ensure patient safety should a component stop operating during a patient examination. Facilities should verify if their system has been impacted by checking if the catalog numbers listed are referenced on the PDB panel located
Quantity378
DistributionNationwide Distribution --- including states of CA, FL, NY, AK, AL, AR, AZ, PA, FL ,CO, SC, CT, IA, DC, DE, GA, HI, IL , IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NJ, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and GUAM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023178INNOVA 4100
Ge Medical Systems, Inc., Nov 26, 2002
K042053INNOVA 4100, INNOVA 3100
Ge Medical Systems, Inc., Aug 26, 2004
K052157INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ
Ge Medical Systems, LLC, Aug 25, 2005
K052412INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ
GE Healthcare, Sep 16, 2005
K061163INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT
Ge Medical Systems, LLC, Jun 7, 2006