ClearanceBrief
Recall Z-0538-2009 ยท KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Depuy Orthopaedics, Inc.

Depuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+9 STD, sterile, Depuy France; REF 130742209. The device is used as a shoulder implant.

DateJan 15, 2009 (initiated Jul 10, 2008)
StatusTerminated
ReasonCompressed air may become trapped between the cup and the stem, preventing the tapered walls from engaging and causing the cup to feel unstable.
Root causeDevice Design
ActionA letter, "Urgent Device Correction" dated July 8, 2008 was sent via certified mail to implanting surgeons advising them of the issues and guidance on how to mitigate the problem. A communication was also sent to all sales representatives and territory offices on July 15, 2008. For questions, please contact Depuy Orthopaedics, Inc. Customer Service at 1-800-473-3789.
Quantity1,521 of all devices.
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062250DELTA XTEND REVERSE SHOULDER SYSTEM
DePuy Orthopaedics, Inc., Feb 2, 2007
K081620MODIFICATION TO DEPUY DELTA XTEND REVERSE SHOULDER SYSTEM
DePuy Orthopaedics, Inc., Jun 30, 2008