ClearanceBrief
Recall Z-0522-2009 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Medical Systems

Philips Diagnostic Ultrasound System, Model HD3. Product labeled HD3 Ultrasound System, Distributed by Philips Ultrasound, 22100 Bothell-Everett Highway, Bothell, WA 98021. Pulse Doppler Imaging Systems used for diagnostic ultrasound imaging or fluid analysis of human body.

DateDec 31, 2008 (initiated Aug 12, 2008)
StatusTerminated
ReasonThe system has a software defect that affects the display of Estimated Fetal Weight (EFW) Growth Percentile result values.
Root causeSoftware design
ActionOn 8/12/08, the firm sent an Urgent - Medical Device Correction letter, dated July 2008, to all affected customers via UPS. The letter describes the reason for the recall, informs them that Philips Ultrasound is correcting the error, states that the software is under development, will be available in September 2008 and will contact the customer to upgrade their system. "In the interim, Philips recommends that you only use the Hadlock or Osaka growth tables to perform your EFW Growth Percentile calculations. You should make the appropriate selections in the Analysis Setup menu on your system.
Quantity4279 units
DistributionWorldwide Distribution --- including USA and countries of Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bosnia, Brazil, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Dominican Rep, Ecuador, Egypt, El Salvador, France, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, In
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031552MYCOLOR 202 DIAGNOSTIC ULTRASOUND SYSTEM AND TRANSDUCERS
Medison America, Inc., May 30, 2003