Recall Z-2349-2008 ยท HTQ Broach

Stryker Howmedica Osteonics Corp.

Rejuvenate Broach Catalog No: 1601-1005, 1601-1006, 1601-1007, 1601-1008, 1601-1009, 1601-1010 Stryker Orthopaedics, Mahwah, NJ. 07430 Contours the proximal femur to the geometry of the stem to provide a complimentary fit of the stem to the bone.The broaches are designed with a series of teeth that are intended to seat against highly densified cancellous bone for the interference press-fit in the

DateSep 19, 2008 (initiated Jan 25, 2008)
StatusTerminated
ReasonRejuvenate Broach system was designed in a manner that may result in the potential for a medical calcar fracture.
Root causeDevice Design
ActionStryker notified all branches and agencies to return the product on January 25, 2008. A formal recall letter was not written as all product was in their control.
Quantity78 devices
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.