Recall Z-2406-2008 ยท DXT Injector And Syringe, Angiographic

Acist Medical Systems

ACIST" BRACCO, AT P54 AngioTouch Kit , SKU # 014644, Sterile EO Manufacturer; ACTB Medical Systems, Inc. 7905 Fuller Road, Eden Prairie, MN 55344, Made in Mexico. Catalogue No: 800608-013 Intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.

DateSep 23, 2008 (initiated Jun 6, 2008)
StatusTerminated
ReasonSome of the sterile package seals of AT P54 AngioTouch Kit were breached. Breached package seals could compromise product sterility, which could potentially lead to patient infection.
Root causePackage design/selection
ActionConsignees were sent a "Urgent: Medical Device Recall" letter dated July 9. 2008. The letter addressed to "Manager" described the Proben and product. The letter advised customers that if they had inventory of any of the lots under the scope of the recall, to stop further distribution of that product and immediately cease use of the product. The letter listed the applicable lots and asked customers to verify their inventory against the list. Any inventories matching the lots on the recall list were to be checked on the provided form and the quantity recorded. They were instructed to return t
Quantity11,750 (USA), 22,230 (OUS)
DistributionWorldwide Distribution --- USA including states AR, AZ, DC, FL, GA, MS, NC, NE, NH, NY, OH, OK, PA, TX, TN,VA, AL, CA, IL, KS, KY, LA, MD, MN, MO, SD, CO, IA, IN, MI, NV, PR, SC, MI, OR, WV, CT, ND, DC, WA, ME, and NJ, and countries of Canada, Brazil, Japan, China, Korea, Thailand, Hong Kong, and Au
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040298ACIST ANTIOGRAPHIC INJECTION SYSTEM CONTRAST MANAGEMENT DISPOSABLE KITS
Acist Medical Systems, Inc., Mar 15, 2004