Recall Z-2400-2008 ยท NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2

DiaDexus, Inc

PLAC Test ELISA kit, catalog number 90106, enzyme immunoassay for the quantitative determination of Lp-PLA2 in human plasma and serum, kit containing coated microwell stripplate and reagents (12 strips, 1 set calibrators 1 to 6, 0.25 ml each, 20x wash buffer, 50 mL, conjugate 23 mL, TMB, 11 mL, Stop solution, 11 mL

DateSep 26, 2008 (initiated Jul 18, 2008)
StatusTerminated
ReasonProduct may give Lp-PLA2 values at up to 22% lower than accurate values.
Root causeFinished device change control
ActionAll customers received communication of product correction dated July 18. 2008, via Federal Express, which included a cover letter, technical bulletin, specimen handling best practices cards, and revised package inserts. E-mail communication was also sent to all but seven customers. Contact diaDexus Technical Support at 1-877-752-2837 for assistance.
Quantity65 (total)
DistributionUS, OUS: UK and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030477DIADEXUS PLAC TEST
Diadexus, Inc., Jul 18, 2003