Recall Z-2399-2008 ยท NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2

DiaDexus, Inc

PLAC Test Reagent Kit, Catalog numbers 90107 and 90110, Turbidimetric Immunoassay for the Quantitative Determination of Lp-PLA2 in human plasma or serum. Product consists of a liquid, ready to use, two reagent kit assembled in a kit box.

DateSep 26, 2008 (initiated Jul 18, 2008)
StatusTerminated
ReasonProduct may give Lp-PLA2 values at up to 22% lower than accurate values.
Root causeFinished device change control
ActionAll customers received communication of product correction dated July 18. 2008, via Federal Express, which included a cover letter, technical bulletin, specimen handling best practices cards, and revised package inserts. E-mail communication was also sent to all but seven customers. Contact diaDexus Technical Support at 1-877-752-2837 for assistance.
Quantity65 (total)
DistributionUS, OUS: UK and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030477DIADEXUS PLAC TEST
Diadexus, Inc., Jul 18, 2003