Recall Z-2395-2008 ยท GEY Motor, Surgical Instrument, Ac-Powered

Synthes U S A

Power Drive. Battery-driven power tool system intended for use during surgical procedures to provide power to operate various accessories and attachments.

DateSep 21, 2008 (initiated Jul 23, 2008)
StatusTerminated
ReasonThe product is mislabeled, as the validation for the previous sterilization parameters cannot be replicated. The currently validated sterilization parameter is a minium of 24 minutes prevacuum.
Root causeError in labeling
ActionAll Synthes Trauma customers and sales consultant s were notified of the correction via return receipt letters sent via UPS. This notification communication contains product literature documenting the validated sterilization parameters. Instructions detail disposal of the previously received literature and replacement with the provided literature. A toll free number with a dedicated extension has been established for questions. The firm has requested that all recipients acknowledge understanding of instructions by returning a Verification Section via fax or email. If you have questions, call
Quantity584 units
DistributionThe product was shipped to medical facilities and sales consultants nationwide. The product was also shipped to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.