Recall Z-2201-2008 ยท CAW Generator, Oxygen, Portable

Pacific Consolidated Industries LLC

Pacific Consolidated Industries Mobile Oxygen Storage Tank, MOST 20-2B, Part Number: 793080-002

DateAug 28, 2008 (initiated Sep 8, 2007)
StatusTerminated
ReasonThe recall was initiated after the firm received two customer complaints of ruptured bourdon tubes that had resulted in bellowed out face gauges. With one unit the operator was able to shut off the oxygen flow, however one rupture resulted in a fire. No injuries were reported.
Root causeVendor change control
ActionPCI sent a Recall Notification letter to the Spanish Air Force September 8, 2007. The Recall Notification letter was mailed to the US Defense Supply Center on July 24, 2008 after a recall agreement was signed on July 8, 2008 between Pacific Consolidated Industries and the Defense Supply Center. The recall notification to Spanish Air Force informed the consignee of the problem and asked the consignee to remove the affected units from service until a Technical Service Representative could be sent to Spain to inspect the affected units and perform the corrective action. The recall notificati
Quantity1032 units
DistributionWorldwide to US Air Force and also to Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040738MOBILE OXYGEN STORAGE TANK (MOST)
Pacific Consolidated Industries Llp, Apr 30, 2004
K061414EXPEDITIONARY DELPLOYABLE OXYGEN CONCENTRATION SYSTEM (EDOCS), MODELS 120B AND 30;MOBILE OXYGEN STORAGE TANK (MOST), MOD
Pacific Consolidated Industries, LLC, Jan 22, 2007