Recall Z-0014-2009 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Intraop Medical Corporation

Mobetron Model 1000, mobile electron linear accelerator, a device system which delivers radiation to an intraoperative site.

DateOct 3, 2008 (initiated Jun 20, 2008)
StatusTerminated
ReasonDocking screw/nut failure, which may result in a treatment head dropping, potentially impacting the patient.
Root causeComponent design/selection
ActionRecall initiated on 06/20/2008, advising users via Advisory Notice that a safety upgrade would be available. Two options are given to either self inspect the product or schedule an Intraop Authorized Field Service Engineer to visit their facility to perform the inspection. The letter notes that if no nut failure was evident, the machine could still be used. The firm continues to send retrofit kits.
Quantity24 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981112MOBETRON
Intraop Medical, Inc., Jul 24, 1998