ClearanceBrief
Recall Z-2379-2008 ยท FQO Table, Operating-Room, Ac-Powered

Stryker Communications Corp

VERTIER Surgical Table, Catalog #0788-100-001S (SIDNE-enabled standard table with 5th wheel); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.

DateSep 18, 2008 (initiated Jul 2, 2008)
StatusTerminated
ReasonHydraulic lines responsible for tilting the surgical table have the potential to be severed when articulating table to its lowest position, possibly resulting in the unexpected and rapid movement of the table.
Root causeDevice Design
ActionUrgent: Device Recall letters were sent to consignees on 07/02/08 via FedEx with return receipt. Consignees were asked to discontinue use of the tables until a Stryker representative has come to the facility and updated the table. The firm has identified and validated replacement parts that prevent this occurrence.
Quantity4 units.
DistributionNationwide Distribution --- including states of AR, CA, FL, GA, ID, IL, IN, KS, KY, MN, MO, ND, NJ, NY, OH, OK, OR, SD, TX, UT, VA and Washington DC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.