Philips Medical Systems

Philips Telemonitoring System Software (M3811B) used with Philips Telemonitoring System (M3810B) Revision: B.02.07 Prescriptive medical device used to automatically collect and transmit medical information (weight, blood pressure, non-diagnostic ECG) over phone lines between provider and patient.

DateSep 18, 2008 (initiated Jul 22, 2008)
StatusTerminated
ReasonSoftware: If Weight Limits edited, values will return to default values that were deleted
Root causeSoftware design
ActionAffected customers will receive the Urgent Medical Device Correction notification on 7/22/08 via certified mail. Delivery receipt will be tracked. The Urgent Medical Device Correction notification instructs customers to verify that daily and weekly weight limits are set or not set as intended while they await the correction, which will come in the form of a new version of software and will simultaneously update all systems. Philips has corrected the software defect and is in the process of validating the correction and releasing the corrected software as part of a scheduled release. If furth
Quantity49 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041674THE M3810A PHILIPS TELEMONITORING SYSTEM WITH M3814A SPO2 UNIT
Philips Medical Systems, Jul 19, 2004