Recall Z-2297-2008 ยท IKX Aid, Transfer
Zimmer Inc.
Zimmer Patient Helper Reinforce Overhead Bar for Hill-Rom Versa Care beds; Zimmer Inc., Warsaw, The product is a L shaped bar that is used with hospital beds set up for traction. The bar allows mounting of a trapeze to help the patient pull themselves up when in the hospital bed. IN; Prod. Nos. 00-2700-965-02 and 00-0965-001-00.
| Date | Sep 16, 2008 (initiated Jul 7, 2008) |
| Status | Terminated |
| Reason | If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar. |
| Root cause | Device Design |
| Action | Consignees were notified via recall letter dated 7/7/08 to discontinue use of these devices until the firm visits and replaces the devices. Contact Zimmer at 1-800-321-5533 for assistance. |
| Quantity | 645 of all systems |
| Distribution | Nationwide and Canada. |