Recall Z-2347-2008 ยท LXH Orthopedic Manual Surgical Instrument

Kyphon Inc

KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First Fracture, 15/3 Xpander), catalog number KPT1502, manufactured by Medtronic Spine LLC. Device is used to access vertebral body during balloon kyphoplasty, a procedure designed to reduce and stabilize vertebral body compression fractures (VCFS).

DateSep 18, 2008 (initiated Jun 23, 2008)
StatusTerminated
ReasonSome products may contain the incorrect introducer stylet. If the treating physician is unfamiliar with the use and design of the diamond tip stylet, there is a possibility of serious injuring occurring with the use of the diamond tip stylet during balloon kyphoplasty procedure. Other risks include delay to surgery and inconvenience to patient.
Root causeMixed-up of materials/components
ActionRecall initiated on June 23, 2008. A recall letter and reconciliation form was sent out via e-mail to sales representatives for hand delivery to US hospital consignees. The consignees were asked to review the letter and complete/return the reconciliation form to the firm. If you have any questions contact Kyphon, Inc. at 1-408-548-6740.
Quantity310 units
DistributionClass II Recall - Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.