Recall Z-1652-2008 ยท DSI Detector And Alarm, Arrhythmia

Spacelabs Healthcare, Incorporated

Ultraview Digital Telemetry Transmitter, Model 90478 The Spacelabs Medical Digital Telemetry system provides continuous electrocardiographic (ECG) monitoring to detect ST segment changes, abnormal cardiac rhythms, including asystole, ventricular fibrillation and ventricular tachycardia.

DateSep 18, 2008 (initiated Jul 18, 2008)
StatusTerminated
ReasonTelemetry Modules are failing to alarm for low heart rate and asystole.
Root causeSoftware design
ActionOn 07/18/08 Spacelabs Healthcare issued a "URGENT-MEDICAL DEVICE CORRECTION" letter, dated July 16, 2008, to all customers who received the Spacelabs Medical Telemetry Receiver-Model 90478. The letter states circumstances in which the alarm fails, how to test each module for high rate alarm prior to patient use and a temporary work around if the alarm failure exists.` Letters have a receipt acknowledgement to ensure each customer receives the letter. Spacelabs Technical Support will personally install the new software, version V3.05.07EN, in all Spacelabs Medical Telemetry Receivers Model
Quantity37,590
DistributionNationwide and Internationally
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K925510MODEL 90478 & 90479 TELEMETRY REC. MODULE&HOUSING
Spacelabs Medical, Inc., Oct 27, 1994