ClearanceBrief
Recall Z-0851-2010 ยท KPA Slide Stainer, Automated

Beckman Coulter Inc

GEN*S SlideMaker, Barcode Label Specifications, GEN*S SM Intended Use, GEN*S Slide and Sample, Part Numbers 6605360, P/N 4237221, 4237212, and 4277299. Intended for use as an optional peripheral to the COULTER GEN*S System. Creates a blood smear on a clean microscope slide using a segment of the blood sample aspirated by the GEN*S System.

DateFeb 25, 2010 (initiated May 29, 2008)
StatusTerminated
ReasonBeckman Coulter has become aware of a potential issue in which the Sample ID information embedded in the barcode affixed to a slide made by the Coulter GEN*S and LH SlideMaker does not match the text.
Root causePending
ActionAn Urgent: Product Corrective Action letter, dated June 9, 2008, was sent to customers. The letter identified the affected products, stated the background, discussed the issue, and stated the actions to be taken. Customers are to: 1. Only use upper case alpha characters in the Sample Identifier 2. If you must use lower case alpha characters in the Sample Identifier, configure the SlideMaker Slide labels without a bar-code. Questions regarding the Product Corrective Action should be directed to Beckman Coulter Customer Service at 1-800-526-7694 or your local Beckman Coulter Rep.
Quantity4
DistributionWorldwide Distribution -- USA and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.