Recall Z-2338-2008 ยท BSZ Gas-Machine, Anesthesia

Draeger Medical, Inc.

Fabius Trio Gas machine, anesthesia or analgesia. Catalog number 8606000. The product is used in operating rooms and ER, for patient intubation.

DateSep 23, 2008 (initiated Jul 7, 2008)
StatusTerminated
ReasonThe parent company became aware of instances where the lower side rail of the frame of the machine broke inwards when being moved across a threshold; resulting in the machine being unstable.
Root causeProcess control
ActionThe firm has issued a MEDICAL DEVICE RECALL letter dated 7/7/08 to its customers via certified mail. The letter informs them of the problem and to exercise care when using the device until it has been retrofitted. Contact Drager Medical at 1-800-543-5047, #4 for assistance.
Quantity111 units
DistributionNationwide