ClearanceBrief
Recall Z-1519-2008 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Datex - Ohmeda, Inc

GE Healthcare Innova 3100 IQ Digital Fluoroscopic Imaging System (Cardiovascular imaging system). The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy of cardiology, dia

DateSep 30, 2008 (initiated Aug 17, 2006)
StatusTerminated
ReasonUser was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to incorrect overall dose management, resulting in unnecessary exposure and can be a contributing factor for skin injury. The alarm can be heard in the control room.
Root causeRadiation Control for Health and Safety Act
ActionGE Field Engineers are visiting all affected customers to apply updated software. For assistance contact Datax-Ohmeda, Inc., at 1-608-221-1551.
Quantity132
Distributionwithin US: AL, AR, AZ, CA,CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico. OUS to include: Algeria, Australia, Belgium, Brazil, Canada, China, Columbia, Denmark, Finland, France, Germany, Greec
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050489INNOVA 2100
Ge Medical Systems, LLC, Mar 11, 2005
K052157INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ
Ge Medical Systems, LLC, Aug 25, 2005
K052412INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ
GE Healthcare, Sep 16, 2005