Recall Z-2336-2008 ยท DSI Detector And Alarm, Arrhythmia

Spacelabs Healthcare, Incorporated

Ultraview Digital Telemetry Transmitter, Model 91341-09 The Spacelabs Medical Digital Telemetry system provides continuous electrocardiographic (ECG) monitoring to detect ST segment changes, abnormal cardiac rhythms, including asystole, ventricular fibrillation and ventricular tachycardia.

DateSep 25, 2008 (initiated Jul 17, 2008)
StatusTerminated
ReasonDuring battery insertion there is potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.
Root causeDevice Design
ActionOn 07/17/08 the firm issued a "URGENT - MEDICAL DEVICE CORRECTION" letter to their customers who received the Spacelabs Medical 1400 MHz Telemetry System - Model 91341-09. The letter states the circumstances in which the device malfunctions, instructions for staff to confirm at the time of the battery insertion that the data is appearing and to remove it from service if information is not appearing. Letters will be sent with a return receipt acknowledging each customer receives the Medical Device Correction letter. Spacelabs Technical Support will respond to all technical questions and pers
Quantity2,055 transmitters
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.