Recall Z-2330-2008 ยท JOJ Indicator, Physical/Chemical Sterilization Process

Propper Mfg Co Inc

Propper short gas-chex EO sterilization indicators, 250 strips per box, Reorder No. 269001, Propper Manufacturing Co., Inc., Long Island City, N.Y. 11101

DateSep 20, 2008 (initiated Feb 7, 2008)
StatusTerminated
ReasonThe firm discovered a production specification discrepancy that may cause the indicators to show an inaccurate result.
Root causeEquipment maintenance
ActionOn 2/07/08, Propper Mfg. Co. sent URGENT: DEVICE RECALL letters (dated 2/07/08) to the direct consignees, informing them of the recall of the affected lots of Gas-Chex EO Indicator Strips. The letters also instructed them to: (1) check their inventory to determine if they have the recalled lots in stock; (2) discontinue use of the affected lots; (3) reprocess any unused packs that have been processed with the affected lots of Gas-Chex EO Indicator Strips with non-recalled lots or an alternative product from another supplier; (4) return the affected lots to Propper Manufacturing; (5) complete t
Quantity1757 cases (7030 boxes)
DistributionNationwide and Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.