Recall Z-2327-2008 ยท CBK Ventilator, Continuous, Facility Use

Draeger Medical, Inc.

Evita XL Critical Care Ventilator. Continuous (Respirator)

DateSep 20, 2008 (initiated Jul 2, 2008)
StatusTerminated
Reasonaudible power failure alarm was not enunciated at the required volume
Root causeNonconforming Material/Component
ActionThe recalling firm issued an URGENT MEDICAL DEVICE RECALL letter dated 7/2/08 informing the consignees of the problem and the need to have the circuit board replaced.
Quantity83 units
DistributionThe products were shipped to medical facilities in AZ, CA, CO, FL, GA, IL, KY, MA, MI, MS, NJ, NY, PA, TN, TX, UT and WA.