ClearanceBrief
Recall Z-2383-2008 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Neuromodulation

Medtronic Intrathecal Catheter Pump Segment Revision Kit, model 8596SC. Contents: 66-cm pump segment with attached sutureless pump connector, spinal segment strain-relief sleeves (2), Pump segment strain-relief sleeves (2), Connector pin.

DateSep 25, 2008 (initiated Jun 26, 2008)
StatusTerminated
ReasonDisconnection of the Medtronic sutureless connector (SC) catheters from the catheter port on the pump, or occlusion between the sutureless pump connector and the catheter port on the pump.
Root causeComponent design/selection
ActionMedtronic sent A Safety Alert dated June 2008, on June 27, 2008, to healthcare professionals, which described the issues of occlusion and disconnect. The letter identifies the affected catheter and revision kit model numbers and the associated implantable infusion pumps. The issues do not involve Medtronic MiniMed insulin pump. The letter refers the HCP to recommendations for Implant Techniques and Recommendations for Patency Verification that are included with the letter. The letter also includes recommendations for managing patients with implanted SC catheters.
Quantity21,674 total
DistributionWorldwide Distribution: USA, Puerto Rico, Algeria, Aruba, Austria, Australia, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Nether
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.