Recall Z-2389-2008 ยท GWQ Full-Montage Standard Electroencephalograph

Cardinal Health NeuroCare Division

NicoletOne photic adapter cable 085-463700

DateSep 21, 2008 (initiated Jun 5, 2008)
StatusTerminated
ReasonCardinal Health NeuroCare is implementing a voluntary field correction to replace a quantity of photic adapter cables, part number 085-463700, used to connect the NicLED Photic Strobe option, part number 698-625500, with NicoletOne EEG system. The photic adapter cable has a wiring error which delivers a reduced voltage and results in a decreased intensity of the NicLED photic. This reduced intensi
Root causeNonconforming Material/Component
ActionA "Cardinal Health Urgent: Medical Device Correction" letter was sent to consignees. The letter described the product and the problem. The consignees were asked to confirm completion of the cable exchange by returning an information card. If you have questions, contact Cardinal Health (formerly VIASYS) NeuroCare Technical Support Group at 1-800-356-0007 (US), 608-273-5000 (OUS) or email shane.sawall@cardinalhealth.com.
Quantity92
DistributionCA FL GA ID IN LA MD ME MI MO NC NY PA TN TX WA WI China Denmark Indonesia Japan Korea Mexico Netherlands Russia Saudi Arabia Slovenia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991054BRAVO MULTI-MODALITY SYSTEM
Nicolet Biomedical, Inc., Nov 3, 1999