Recall Z-2388-2008 ยท MKJ Automated External Defibrillators (Non-Wearable)

Physio Control, Inc.

Physio-Control Lifepak 20 defibrillator/monitor is intended for the termination of certain fatal arrhythmias.

DateSep 21, 2008 (initiated Jun 18, 2008)
StatusTerminated
ReasonPotential for the coin battery to drain prematurely causing the monitor clock time and date to be incorrect and the service light indicator to illuminate.
Root causeNonconforming Material/Component
ActionOn 06/18/08 the firm, via regular postal service , mailed an "IMPORTANT MEDICAL DEVICE NOTIFICATION" letter to all their consignees. The letter instructs consignees that the coin battery may deplete prematurely and to contact Physio-Control if their Lifepack 20 defibrillator/monitor requires coin battery replacement prior to the recommended four year replacement schedule. If you have any questions, contact Technical Support at 1-800-442-1142, option 5 from 6:00 am to 4:00 pm (Pacific), Monday-Friday or visit www.physio-control-notices.com/coin battery.
Quantity45,209 devices
DistributionThere were 24,238 devices distributed to 3,135 consignees throughout the United States and 20,613 devices distributed to 48 foreign consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012274MEDTRONIC PHYSIO-CONTROL LIFEPAK 20 DEFIBRILLATOR/MONITOR/PACEMAKER
Medtronic Physio-Control Corp., Feb 5, 2002