ClearanceBrief
Recall Z-2316-2008 ยท FSA Lift, Patient, Non-Ac-Powered

Liko AB

Universal SlingBar 450 Product Number: 3156075 Used with the: Viking XS, Viking S, Viking M, Viking L Patient Lifts

DateSep 20, 2008 (initiated Jul 1, 2008)
StatusTerminated
ReasonA component securing the Universal SlingBar to the patient lift, may experience a nut unthreading resulting in the sling bar detaching from the lift causing the patient to fall.
Root causeComponent design/selection
ActionLiko North America ,the US Distributor, notified accounts by letter dated June 30, 2008 and provided a Field Safety Notice with Safety Inspection Procedures to performed on the lift prior to use. Replacement slingbars are being provided. If you have questions, contact the Liko product distributor in your country. Contact information can be found at www.liko.com/int/distributors.asp.
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.