Recall Z-2314-2008 ยท DSI Detector And Alarm, Arrhythmia

Spacelabs Healthcare, Incorporated

Spacelabs Medical Temperature Probe Adaptor PN-700-0031-00. Device is packaged in clear plastic bags with an attached insert card Insert card reads "TruLink series Dual Temperature Cable REF 700-0032-00 Spacelabs Medical Issaquah, WA USA"

DateSep 20, 2008 (initiated Jul 2, 2008)
StatusTerminated
ReasonIntermittent contact of temperature probe adaptors can cause a loss of temperature monitoring which can cause a delay in care when an alarm is missed.
Root causeNonconforming Material/Component
ActionStarting July 3, 2008 the firm issued a "MEDICAL DEVICE CORRECTION" letter notifying their customers of the potential hazards with the use of probe adapter, PN 700-031-00. All letters have a return receipt to verify each customer received the medical device correction letter. The firm will replace all PN 700-0031-00 adaptors with a replacement adaptor from their initial distributor. The firm initiated sending new adaptors separately, by mail, with a copy of the same Medical Device Correction letter to the firm's customers. The firm has requested their customers to throw away any adapter t
Quantity3,733
DistributionNationwide and Worldwide to medical facilities.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K972502SPACELABS MEDICAL INTEGRATED MULTIPARAMETER MODULE 90496
Spacelabs Medical, Inc., May 28, 1998