Recall Z-2312-2008 ยท LLZ System, Image Processing, Radiological

GE Healthcare

Advantage Workstation version 4.2P, 4.3 and 4.4 with Volume Viewer 2 and Volume Viewer 3. Model Numbers 5183652 & 5142878.

DateSep 20, 2008 (initiated May 1, 2008)
StatusTerminated
ReasonGE Healthcare has recently became aware of a possible mismatch between the label of the tracked vessel and the underlying image associated with the cardio- vascular applications of the Advantage workstation that may impact patient safety.
Root causeSoftware design
ActionConsignees were sent a GE "Urgent Field Safety Notice" letter dated March 13, 2008 addressed to Hospital Administrators; Managers of Radiology/Cardiology; Radiologists/Cardiologists on May 2008. The letter described the safety issue, affected product, product correction, safety instructions and contact information. A GE Field Engineer to visit site to implement the correction (FMI No. 25409). Contact information: 800-437-1171 (USA); 0120-055-919 (Japan); 0800-15-25-25 (France); for all other countries contact your local GE Healthcare sales or field representative.
Quantity5,249 (2,104 US/ 3,145 OUS)
DistributionAK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041521VOLUME VIEWER PLUS
General Electric Medical Systems, Jun 22, 2004