ClearanceBrief
Recall Z-2287-2008 ยท JLB Atp And Creatine Kinase (Enzymatic), Creatine

Abaxis Inc

Piccolo Chemistry Analyzer software version 6.121 with Electrolyte Panel Lot #8041AC5, model number 400-1022 in polyfoil pouch. Piccolo Electrolyte Panel is a single-use, self-contained reagent disk intended for use with the Piccolo and PiccoloXpress chemistry analyzers, for in vitro qualitative determination.

DateSep 17, 2008 (initiated Apr 21, 2008)
StatusTerminated
ReasonThe product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples
Root causeProcess control
ActionThe firm issued a notification letter to its consignees on April 21, 2008. The firm intends to follow up with phone calls and provide a software upgrade for the products. Contact Abaxis Customer Service at 1-510-675-6500 for assistance.
Quantity237 kits
DistributionProduct was distributed Worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992140PICCOLO METLYTE 7 REAGENT DISC
Abaxis, Inc., Aug 26, 1999