Recall Z-2310-2008 ยท LZS Excimer Laser System

VISX INCORPORATED, A SUBSIDIARY OF AMO INC

AMO WaveScan WaveFront System Version 3.90/3.901, Part Number 0070-1478. The device is used in conjunction with the VISX STAR S4 and STAR S4 IR Excimer Laser Systems, which are indicated for wavefront-guided laser assisted in situ keratomileusis (LASIK) eye surgery in certain patient populations. The WaveScan System is a diagnostic instrument intended for the automated measurement, analysis and

DateSep 17, 2008 (initiated Jun 9, 2008)
StatusTerminated
ReasonInaccurate measurements may be generated by the device, which could result in improper treatment and deterioration of patient eyesight.
Root causeSoftware design
ActionA product advisory notice was mailed to all consignees on June 9, 2008, advising consignees to closely monitor the WaveScan results to ensure that mistreatment is not given to patients. The firm has ceased distributing the updated software. ***UPDATE 08/29/2008*** The firm revised its recall strategy on 08/28/2008 because the firm was informed that a 180 day PMA supplement rather than a real time review supplement would be required for the software upgrade approval. The firm will instead revert products to the previous software version, 3.67/3.671. If you have questions, contact 1-877-AMO-
Quantity160 kits
DistributionProduct was distributed to 152 customers throughout the United States.